Technical Information - Precision
Technical Information - Accuracy
Accuracy
Table 1 – System accuracy results for glucose
concentration < 5.55 mmol/L
Difference range in values
between YSI laboratory
reference method and
CONTOUR NEXT meter
Within ±
0.28
mmol/L
Within ±
0.56
mmol/L
Within ±
0.83
mmol/L
Number (and percent)
of samples within
specifiedrange
167 of 180
(92.8%)
179 of 180
(99.4%)
180 of 180
(100%)
Table 2 – System accuracy results for glucose
concentration ≥ 5.55 mmol/L
Difference range in values
between YSI laboratory
reference method and
CONTOUR NEXT meter
Within ±
5%
Within ±
10%
Within ±
15%
Number (and percent)
of samples within
specifiedrange
319 of 420
(76.0%)
412 of 420
(98.1%)
419 of 420
(99.8%)
A repeatability study was conducted with the CONTOUR
®
NEXT
blood glucose monitoring system using 5 venous
whole blood specimens with glucose levels from 2.2 to
17.9mmol/L. Multiple replicates (n=300) were tested using
multiple CONTOUR NEXT blood glucose meters and 3 lots
of CONTOUR
®
NEXT blood glucose test strips. The following
precision results were obtained.
Table 4 – System repeatability results for C
ONTOUR NEXT
meter using C
ONTOUR NEXT test strips
Mean,
mmol/L
Pooled
Standard
Deviation
95% Confidence
Interval of Standard
Deviation, mmol/L
Coefficient of
Variation, %
2.2 0.05 0.043 – 0.051 2.1
4.8 0.07 0.062 – 0.073 1.4
6.8 0.09 0.082 – 0.097 1.3
11.4 0.15 0.138 – 0.164 1.3
18.2 0.26 0.238 – 0.282 1.4
Intermediate precision (which includes variability across
multiple days) was evaluated using 3 control solutions. With
each of 3 lots of C
ONTOUR NEXT Test Strips, each control
was tested once on each of 10 instruments on 10 separate
days for a total of 300 readings.
Table 5 – System intermediate precision results for
C
ONTOUR NEXT meter using CONTOUR NEXT test strips
Control
Level
Mean,
mmol/L
Standard
Deviation,
mmol/L
95% Confidence
Interval of Standard
Deviation, mmol/L
Coefficient
of Variation,
%
Low 2.41 0.04 0.036 – 0.043 1.6
Normal 7.16 0.10 0.093 – 0.111 1.4
High 21.30 0.35 0.321 – 0.380 1.6
Acceptance criteria in ISO 15197: 2013 are that 95% of all
differences in glucose values (i.e., between reference method and
meter) should be within ±0.83 mmol/L for glucose values less than
5.55 mmol/L, and within ±15% for glucose values greater than or
equal to 5.55 mmol/L.
Technical Information
User Accuracy
The CONTOUR
®
NEXT blood glucose monitoring system
was tested in 100 capillary blood samples using 600
CONTOUR
®
NEXT test strips. Two replicates were tested
with each of 3 lots of CONTOUR NEXT test strips for a
total of 600 readings. Results were compared to the YSI
®
glucose analyzer, which is traceable to the CDC hexokinase
method.
The tables below compare the performance of the
2methods.
Table 3 – System accuracy results for glucose
concentrations between 1.3 mmol/L and 30.6mmol/L
Within ± 0.83 mmol/L or ±15% 599 of 600 (99.8%)
Precision
A study evaluating glucose values from fingertip capillary
blood samples obtained by 221 lay persons showed
the following results: 100% within ± 0.83 mmol/L of the
medical laboratory values at glucose concentrations below
5.55mmol/L and 98.9% within ±15% of the medical
laboratory glucose concentrations at or above 5.55mmol/L.